Design of Transcranial Magnetic Stimulation Coils with Optimal Trade-Off Between Depth, Focality, and Energy

TITLE
Design of Transcranial Magnetic Stimulation Coils with Optimal Trade-Off Between Depth, Focality, and Energy

SOURCE
Journal of Neural Engineering. 15(4):046033, 2018 08.

AUTHORS
Gomez LJ; Goetz SM; Peterchev AV. Institution Gomez, Luis J. Department of Psychiatry and Behavioral Sciences, Duke University, Durham, NC 27710, United States of America.

OBJECTIVE
Transcranial magnetic stimulation (TMS) is a noninvasive brain stimulation technique used for research and clinical applications. Existent TMS coils are limited in their precision of spatial targeting (focality), especially for deeper targets. This paper presents a methodology for designing TMS coils to achieve optimal trade-off between the depth and focality of the induced electric field (E-field), as well as the energy required by the coil.

APPROACH
A multi-objective optimization technique is used for computationally designing TMS coils that achieve optimal trade-offs between E-field focality, depth, and energy (fdTMS coils). The fdTMS coil winding(s) maximize focality (minimize the volume of the brain region with E-field above a given threshold) while reaching a target at a specified depth and not exceeding predefined peak E-field strength and required coil energy. Spherical and MRI-derived head models are used to compute the fundamental depth-focality trade-off as well as focality-energy trade-offs for specific target depths.

MAIN RESULTS
Across stimulation target depths of 1.0-3.4 cm from the brain surface, the suprathreshold volume can be theoretically decreased by 42%-55% compared to existing TMS coil designs. The suprathreshold volume of a figure-8 coil can be decreased by 36%, 44%, or 46%, for matched, doubled, or quadrupled energy. For matched focality and energy, the depth of a figure-8 coil can be increased by 22%.

SIGNIFICANCE
Computational design of TMS coils could enable more selective targeting of the induced E-field. The presented results appear to be the first significant advancement in the depth-focality trade-off of TMS coils since the introduction of the figure-8 coil three decades ago, and likely represent the fundamental physical limit.